Approval Required

No, but must be collected at Parnassus location. Contact Hematology Lab at 415-353-1747 to receive special collection kit. For Neurointerventional radiology only.

Available Stat

No

Performing Lab

Hematology, Parnassus

Performed

Monday - Friday 0800 - 1600
Saturday - Sunday 0800 - 1545

Methodology

VerifyNow System

Reported

4 hours

Additional Information

The VerifyNow Aspirin Test is designed to measure aspirin's expected antiplatelet effect.

Plavix®, Ticlid®, and Effient® are commonly prescribed in conjunction with aspirin. While infrequent, these agents may cause a reduction of ARU in some patients.

Test performance was not affected by hematocrit values between 29-56%, platelet count values of >= 92,000/µL or moderate to extensive blood hemolysis induced by physical manipulation. Based on an in-house study performed on 3/29/16, correlation showed that as platelet count increased, ARU results increased.

According to manufacturer:
≥ 550 ARU - Platelet dysfunction consistent with aspirin has not been detected

< 550 ARU - Platelet dysfunction consistent with aspirin has been detected

Results should be interpreted in conjunction with other laboratory and clinical data available to the clinician.

Synonyms

  • Aspirin Reaction

Sample Type

Citrated and EDTA anticoagulated whole blood

Collect

1. Greiner partial fill discard
2. Greiner partial fill 3.2% Sodium citrate blue top
3. Lavender EDTA tube

Amount to Collect

8 mL

Preferred Volume

8 mL

Minimum Volume

6 mL

Remarks

Contact Hematology at 415-353-1747 for collection kits.  DO NOT collect any specimens before 0800 or after 1500 from Monday to Friday; before 0800 or after 1500 on Saturday and Sunday.
 
1. Whole blood samples must be collected in or immediately transferred to Greiner 2.0 mL partial fill blue top tubes containing 3.2% Sodium Citrate. The tube must be filled to its intended whole blood capacity (indicated by small black line).
2. Whole blood may be collected from venous sites using a 21 gauge or larger needle in an appropriate blood collection tube.
3. Blood samples should be obtained from an extremity free of peripheral venous infusions.
4. Collect a discard tube first (approximately 2 mL), Greiner partial fill blue tops 2nd, and a lavender for CBC last.
5. Gently invert the sample tube at least 5 times to ensure complete mixing of the contents.
6. Blood must set a minimum of 30 minutes after collection before testing but no longer than 4 hours.

 Samples cannot be pneumatic tubed.
 

Stability (from collection to initiation)

4 hours

Unacceptable Conditions

Clotted samples or if stability period exceeded.

Incorrect tube type and/or no discard tube received.

Patients with Hct's < 33% and/or Plt's < 119

Test Code

ARU

Performing Lab

Hematology, Parnassus

Specimen Preparation

Deliver immediately to Hematology

Additional Processing Instructions

An ARU will be ordered in computer. If no CBC or CBCD is requested, Central Processing will order a hematocrit (HCT) and platelet count (PLT).

Preferred Volume

8 mL

Minimum Volume

6 mL

Unacceptable Conditions

Clotted samples or if stability period exceeded.

Incorrect tube type and/or no discard tube received.

Patients with Hct's < 33% and/or Plt's < 119

Stability (from collection to initiation)

4 hours

Units

ARU

Reference Interval

620 - 672

Additional Information

The VerifyNow Aspirin Test is designed to measure aspirin's expected antiplatelet effect.

Plavix®, Ticlid®, and Effient® are commonly prescribed in conjunction with aspirin. While infrequent, these agents may cause a reduction of ARU in some patients.

Test performance was not affected by hematocrit values between 29-56%, platelet count values of >= 92,000/µL or moderate to extensive blood hemolysis induced by physical manipulation. Based on an in-house study performed on 3/29/16, correlation showed that as platelet count increased, ARU results increased.

According to manufacturer:
≥ 550 ARU - Platelet dysfunction consistent with aspirin has not been detected

< 550 ARU - Platelet dysfunction consistent with aspirin has been detected

Results should be interpreted in conjunction with other laboratory and clinical data available to the clinician.

CPT Codes

85576

LOINC Codes

49011-0

Approval Required

No, but must be collected at Parnassus location. Contact Hematology Lab at 415-353-1747 to receive special collection kit. For Neurointerventional radiology only.

Available Stat

No

Test Code

ARU

Performing Lab

Hematology, Parnassus

Performed

Monday - Friday 0800 - 1600
Saturday - Sunday 0800 - 1545

Methodology

VerifyNow System

Remarks

Contact Hematology at 415-353-1747 for collection kits.  DO NOT collect any specimens before 0800 or after 1500 from Monday to Friday; before 0800 or after 1500 on Saturday and Sunday.
 
1. Whole blood samples must be collected in or immediately transferred to Greiner 2.0 mL partial fill blue top tubes containing 3.2% Sodium Citrate. The tube must be filled to its intended whole blood capacity (indicated by small black line).
2. Whole blood may be collected from venous sites using a 21 gauge or larger needle in an appropriate blood collection tube.
3. Blood samples should be obtained from an extremity free of peripheral venous infusions.
4. Collect a discard tube first (approximately 2 mL), Greiner partial fill blue tops 2nd, and a lavender for CBC last.
5. Gently invert the sample tube at least 5 times to ensure complete mixing of the contents.
6. Blood must set a minimum of 30 minutes after collection before testing but no longer than 4 hours.

 Samples cannot be pneumatic tubed.
 

Collect

1. Greiner partial fill discard
2. Greiner partial fill 3.2% Sodium citrate blue top
3. Lavender EDTA tube

Amount to Collect

8 mL

Sample Type

Citrated and EDTA anticoagulated whole blood

Preferred Volume

8 mL

Minimum Volume

6 mL

Unacceptable Conditions

Clotted samples or if stability period exceeded.

Incorrect tube type and/or no discard tube received.

Patients with Hct's < 33% and/or Plt's < 119

Specimen Preparation

Deliver immediately to Hematology

Additional Processing Instructions

An ARU will be ordered in computer. If no CBC or CBCD is requested, Central Processing will order a hematocrit (HCT) and platelet count (PLT).

Units

ARU

Reference Interval

620 - 672

Synonyms

  • Aspirin Reaction

Stability (from collection to initiation)

4 hours

Reported

4 hours

Additional Information

The VerifyNow Aspirin Test is designed to measure aspirin's expected antiplatelet effect.

Plavix®, Ticlid®, and Effient® are commonly prescribed in conjunction with aspirin. While infrequent, these agents may cause a reduction of ARU in some patients.

Test performance was not affected by hematocrit values between 29-56%, platelet count values of >= 92,000/µL or moderate to extensive blood hemolysis induced by physical manipulation. Based on an in-house study performed on 3/29/16, correlation showed that as platelet count increased, ARU results increased.

According to manufacturer:
≥ 550 ARU - Platelet dysfunction consistent with aspirin has not been detected

< 550 ARU - Platelet dysfunction consistent with aspirin has been detected

Results should be interpreted in conjunction with other laboratory and clinical data available to the clinician.

CPT Codes

85576

LOINC Codes

49011-0
Ordering

Approval Required

No, but must be collected at Parnassus location. Contact Hematology Lab at 415-353-1747 to receive special collection kit. For Neurointerventional radiology only.

Available Stat

No

Performing Lab

Hematology, Parnassus

Performed

Monday - Friday 0800 - 1600
Saturday - Sunday 0800 - 1545

Methodology

VerifyNow System

Reported

4 hours

Additional Information

The VerifyNow Aspirin Test is designed to measure aspirin's expected antiplatelet effect.

Plavix®, Ticlid®, and Effient® are commonly prescribed in conjunction with aspirin. While infrequent, these agents may cause a reduction of ARU in some patients.

Test performance was not affected by hematocrit values between 29-56%, platelet count values of >= 92,000/µL or moderate to extensive blood hemolysis induced by physical manipulation. Based on an in-house study performed on 3/29/16, correlation showed that as platelet count increased, ARU results increased.

According to manufacturer:
≥ 550 ARU - Platelet dysfunction consistent with aspirin has not been detected

< 550 ARU - Platelet dysfunction consistent with aspirin has been detected

Results should be interpreted in conjunction with other laboratory and clinical data available to the clinician.

Synonyms

  • Aspirin Reaction
Collection

Sample Type

Citrated and EDTA anticoagulated whole blood

Collect

1. Greiner partial fill discard
2. Greiner partial fill 3.2% Sodium citrate blue top
3. Lavender EDTA tube

Amount to Collect

8 mL

Preferred Volume

8 mL

Minimum Volume

6 mL

Remarks

Contact Hematology at 415-353-1747 for collection kits.  DO NOT collect any specimens before 0800 or after 1500 from Monday to Friday; before 0800 or after 1500 on Saturday and Sunday.
 
1. Whole blood samples must be collected in or immediately transferred to Greiner 2.0 mL partial fill blue top tubes containing 3.2% Sodium Citrate. The tube must be filled to its intended whole blood capacity (indicated by small black line).
2. Whole blood may be collected from venous sites using a 21 gauge or larger needle in an appropriate blood collection tube.
3. Blood samples should be obtained from an extremity free of peripheral venous infusions.
4. Collect a discard tube first (approximately 2 mL), Greiner partial fill blue tops 2nd, and a lavender for CBC last.
5. Gently invert the sample tube at least 5 times to ensure complete mixing of the contents.
6. Blood must set a minimum of 30 minutes after collection before testing but no longer than 4 hours.

 Samples cannot be pneumatic tubed.
 

Stability (from collection to initiation)

4 hours

Unacceptable Conditions

Clotted samples or if stability period exceeded.

Incorrect tube type and/or no discard tube received.

Patients with Hct's < 33% and/or Plt's < 119
Processing

Test Code

ARU

Performing Lab

Hematology, Parnassus

Specimen Preparation

Deliver immediately to Hematology

Additional Processing Instructions

An ARU will be ordered in computer. If no CBC or CBCD is requested, Central Processing will order a hematocrit (HCT) and platelet count (PLT).

Preferred Volume

8 mL

Minimum Volume

6 mL

Unacceptable Conditions

Clotted samples or if stability period exceeded.

Incorrect tube type and/or no discard tube received.

Patients with Hct's < 33% and/or Plt's < 119

Stability (from collection to initiation)

4 hours
Result Interpretation

Units

ARU

Reference Interval

620 - 672

Additional Information

The VerifyNow Aspirin Test is designed to measure aspirin's expected antiplatelet effect.

Plavix®, Ticlid®, and Effient® are commonly prescribed in conjunction with aspirin. While infrequent, these agents may cause a reduction of ARU in some patients.

Test performance was not affected by hematocrit values between 29-56%, platelet count values of >= 92,000/µL or moderate to extensive blood hemolysis induced by physical manipulation. Based on an in-house study performed on 3/29/16, correlation showed that as platelet count increased, ARU results increased.

According to manufacturer:
≥ 550 ARU - Platelet dysfunction consistent with aspirin has not been detected

< 550 ARU - Platelet dysfunction consistent with aspirin has been detected

Results should be interpreted in conjunction with other laboratory and clinical data available to the clinician.
Administrative

CPT Codes

85576

LOINC Codes

49011-0
Complete View

Approval Required

No, but must be collected at Parnassus location. Contact Hematology Lab at 415-353-1747 to receive special collection kit. For Neurointerventional radiology only.

Available Stat

No

Test Code

ARU

Performing Lab

Hematology, Parnassus

Performed

Monday - Friday 0800 - 1600
Saturday - Sunday 0800 - 1545

Methodology

VerifyNow System

Remarks

Contact Hematology at 415-353-1747 for collection kits.  DO NOT collect any specimens before 0800 or after 1500 from Monday to Friday; before 0800 or after 1500 on Saturday and Sunday.
 
1. Whole blood samples must be collected in or immediately transferred to Greiner 2.0 mL partial fill blue top tubes containing 3.2% Sodium Citrate. The tube must be filled to its intended whole blood capacity (indicated by small black line).
2. Whole blood may be collected from venous sites using a 21 gauge or larger needle in an appropriate blood collection tube.
3. Blood samples should be obtained from an extremity free of peripheral venous infusions.
4. Collect a discard tube first (approximately 2 mL), Greiner partial fill blue tops 2nd, and a lavender for CBC last.
5. Gently invert the sample tube at least 5 times to ensure complete mixing of the contents.
6. Blood must set a minimum of 30 minutes after collection before testing but no longer than 4 hours.

 Samples cannot be pneumatic tubed.
 

Collect

1. Greiner partial fill discard
2. Greiner partial fill 3.2% Sodium citrate blue top
3. Lavender EDTA tube

Amount to Collect

8 mL

Sample Type

Citrated and EDTA anticoagulated whole blood

Preferred Volume

8 mL

Minimum Volume

6 mL

Unacceptable Conditions

Clotted samples or if stability period exceeded.

Incorrect tube type and/or no discard tube received.

Patients with Hct's < 33% and/or Plt's < 119

Specimen Preparation

Deliver immediately to Hematology

Additional Processing Instructions

An ARU will be ordered in computer. If no CBC or CBCD is requested, Central Processing will order a hematocrit (HCT) and platelet count (PLT).

Units

ARU

Reference Interval

620 - 672

Synonyms

  • Aspirin Reaction

Stability (from collection to initiation)

4 hours

Reported

4 hours

Additional Information

The VerifyNow Aspirin Test is designed to measure aspirin's expected antiplatelet effect.

Plavix®, Ticlid®, and Effient® are commonly prescribed in conjunction with aspirin. While infrequent, these agents may cause a reduction of ARU in some patients.

Test performance was not affected by hematocrit values between 29-56%, platelet count values of >= 92,000/µL or moderate to extensive blood hemolysis induced by physical manipulation. Based on an in-house study performed on 3/29/16, correlation showed that as platelet count increased, ARU results increased.

According to manufacturer:
≥ 550 ARU - Platelet dysfunction consistent with aspirin has not been detected

< 550 ARU - Platelet dysfunction consistent with aspirin has been detected

Results should be interpreted in conjunction with other laboratory and clinical data available to the clinician.

CPT Codes

85576

LOINC Codes

49011-0