Available Stat

No

Performing Lab

China Basin Chemistry

Performed

Monday - Friday (day shift)

Methodology

Chemiluminescent Microparticle Immunoassay (Abbott Architect i2000)

Reported

1-4 days

Additional Information

Method changed from Siemens Centaur XP platform to Abbott Architect i2000 platform on X3/19/2018. The Abbott Architect method reads approximately 30% higher than the Centaur method.  No changes were made to the reference range.

Results for this test are generated by immunoassay using the Abbott Architect i2000 platform.  Results for this test determined by assay methods from other manufacturers may not be comparable.

The ARCHITECT CA 125 II Calibrators are manufactured volumetrically and are referenced to this standard prepared by Fujirebio Diagnostics, Inc. There is no internationally recognized CA 125 standard available at this time.

Synonyms

  • CA125
  • C125

Sample Type

Serum

Collect

Gold or Red top preferred. Dark green or light green acceptable.

Amount to Collect

1 mL blood

Preferred Volume

0.3 mL serum

Minimum Volume

0.1 mL serum

Stability (from collection to initiation)

Refrigerated (2-8°C):  7 days
Frozen (-20°C or colder): 12 months

If testing will be delayed more than 24 hours, remove serum from clot, red blood cells or serum separator gel.

Avoid multiple freeze-thaw cycles.

Storage/Transport Temperature

-20°C or colder

Test Code

C125

Performing Lab

China Basin Chemistry

Specimen Preparation

Aliquot and refrigerate serum at 2-8°C.

Preferred Volume

0.3 mL serum

Minimum Volume

0.1 mL serum

Stability (from collection to initiation)

Refrigerated (2-8°C):  7 days
Frozen (-20°C or colder): 12 months

If testing will be delayed more than 24 hours, remove serum from clot, red blood cells or serum separator gel.

Avoid multiple freeze-thaw cycles.

Storage/Transport Temperature

-20°C or colder

Units

U/mL

Reference Interval

Adult Female Reference Range
Age Reference Range (U/mL)
≥ 18 years < 36

Reference range adopted from Abbott (vendor) based on in-house verification study of 20 normal female normal volunteers (>18 years old) in the UCSF Laboratory.

Pediatric Female Reference Range
Age Reference Range (U/mL)
0 - < 4 months 2 - 22
4 months - < 5 years 6-39
5 years - < 11 years 5 - 30
11 years - < 18 years 8-33

Pediatric reference ranges adopted from CALIPER Pediatric Reference Interval study performed on random samples using the Abbott Architect i2000 assay.

Reference range is not established for males.

Additional Information

Method changed from Siemens Centaur XP platform to Abbott Architect i2000 platform on X3/19/2018. The Abbott Architect method reads approximately 30% higher than the Centaur method.  No changes were made to the reference range.

Results for this test are generated by immunoassay using the Abbott Architect i2000 platform.  Results for this test determined by assay methods from other manufacturers may not be comparable.

The ARCHITECT CA 125 II Calibrators are manufactured volumetrically and are referenced to this standard prepared by Fujirebio Diagnostics, Inc. There is no internationally recognized CA 125 standard available at this time.

CPT Codes

86304

LOINC Codes

10334-1

Available Stat

No

Test Code

C125

Performing Lab

China Basin Chemistry

Performed

Monday - Friday (day shift)

Methodology

Chemiluminescent Microparticle Immunoassay (Abbott Architect i2000)

Collect

Gold or Red top preferred. Dark green or light green acceptable.

Amount to Collect

1 mL blood

Sample Type

Serum

Preferred Volume

0.3 mL serum

Minimum Volume

0.1 mL serum

Specimen Preparation

Aliquot and refrigerate serum at 2-8°C.

Units

U/mL

Reference Interval

Adult Female Reference Range
Age Reference Range (U/mL)
≥ 18 years < 36

Reference range adopted from Abbott (vendor) based on in-house verification study of 20 normal female normal volunteers (>18 years old) in the UCSF Laboratory.

Pediatric Female Reference Range
Age Reference Range (U/mL)
0 - < 4 months 2 - 22
4 months - < 5 years 6-39
5 years - < 11 years 5 - 30
11 years - < 18 years 8-33

Pediatric reference ranges adopted from CALIPER Pediatric Reference Interval study performed on random samples using the Abbott Architect i2000 assay.

Reference range is not established for males.

Synonyms

  • CA125
  • C125

Storage/Transport Temperature

-20°C or colder

Stability (from collection to initiation)

Refrigerated (2-8°C):  7 days
Frozen (-20°C or colder): 12 months

If testing will be delayed more than 24 hours, remove serum from clot, red blood cells or serum separator gel.

Avoid multiple freeze-thaw cycles.

Reported

1-4 days

Additional Information

Method changed from Siemens Centaur XP platform to Abbott Architect i2000 platform on X3/19/2018. The Abbott Architect method reads approximately 30% higher than the Centaur method.  No changes were made to the reference range.

Results for this test are generated by immunoassay using the Abbott Architect i2000 platform.  Results for this test determined by assay methods from other manufacturers may not be comparable.

The ARCHITECT CA 125 II Calibrators are manufactured volumetrically and are referenced to this standard prepared by Fujirebio Diagnostics, Inc. There is no internationally recognized CA 125 standard available at this time.

CPT Codes

86304

LOINC Codes

10334-1
Ordering

Available Stat

No

Performing Lab

China Basin Chemistry

Performed

Monday - Friday (day shift)

Methodology

Chemiluminescent Microparticle Immunoassay (Abbott Architect i2000)

Reported

1-4 days

Additional Information

Method changed from Siemens Centaur XP platform to Abbott Architect i2000 platform on X3/19/2018. The Abbott Architect method reads approximately 30% higher than the Centaur method.  No changes were made to the reference range.

Results for this test are generated by immunoassay using the Abbott Architect i2000 platform.  Results for this test determined by assay methods from other manufacturers may not be comparable.

The ARCHITECT CA 125 II Calibrators are manufactured volumetrically and are referenced to this standard prepared by Fujirebio Diagnostics, Inc. There is no internationally recognized CA 125 standard available at this time.

Synonyms

  • CA125
  • C125
Collection

Sample Type

Serum

Collect

Gold or Red top preferred. Dark green or light green acceptable.

Amount to Collect

1 mL blood

Preferred Volume

0.3 mL serum

Minimum Volume

0.1 mL serum

Stability (from collection to initiation)

Refrigerated (2-8°C):  7 days
Frozen (-20°C or colder): 12 months

If testing will be delayed more than 24 hours, remove serum from clot, red blood cells or serum separator gel.

Avoid multiple freeze-thaw cycles.

Storage/Transport Temperature

-20°C or colder
Processing

Test Code

C125

Performing Lab

China Basin Chemistry

Specimen Preparation

Aliquot and refrigerate serum at 2-8°C.

Preferred Volume

0.3 mL serum

Minimum Volume

0.1 mL serum

Stability (from collection to initiation)

Refrigerated (2-8°C):  7 days
Frozen (-20°C or colder): 12 months

If testing will be delayed more than 24 hours, remove serum from clot, red blood cells or serum separator gel.

Avoid multiple freeze-thaw cycles.

Storage/Transport Temperature

-20°C or colder
Result Interpretation

Units

U/mL

Reference Interval

Adult Female Reference Range
Age Reference Range (U/mL)
≥ 18 years < 36

Reference range adopted from Abbott (vendor) based on in-house verification study of 20 normal female normal volunteers (>18 years old) in the UCSF Laboratory.

Pediatric Female Reference Range
Age Reference Range (U/mL)
0 - < 4 months 2 - 22
4 months - < 5 years 6-39
5 years - < 11 years 5 - 30
11 years - < 18 years 8-33

Pediatric reference ranges adopted from CALIPER Pediatric Reference Interval study performed on random samples using the Abbott Architect i2000 assay.

Reference range is not established for males.

Additional Information

Method changed from Siemens Centaur XP platform to Abbott Architect i2000 platform on X3/19/2018. The Abbott Architect method reads approximately 30% higher than the Centaur method.  No changes were made to the reference range.

Results for this test are generated by immunoassay using the Abbott Architect i2000 platform.  Results for this test determined by assay methods from other manufacturers may not be comparable.

The ARCHITECT CA 125 II Calibrators are manufactured volumetrically and are referenced to this standard prepared by Fujirebio Diagnostics, Inc. There is no internationally recognized CA 125 standard available at this time.
Administrative

CPT Codes

86304

LOINC Codes

10334-1
Complete View

Available Stat

No

Test Code

C125

Performing Lab

China Basin Chemistry

Performed

Monday - Friday (day shift)

Methodology

Chemiluminescent Microparticle Immunoassay (Abbott Architect i2000)

Collect

Gold or Red top preferred. Dark green or light green acceptable.

Amount to Collect

1 mL blood

Sample Type

Serum

Preferred Volume

0.3 mL serum

Minimum Volume

0.1 mL serum

Specimen Preparation

Aliquot and refrigerate serum at 2-8°C.

Units

U/mL

Reference Interval

Adult Female Reference Range
Age Reference Range (U/mL)
≥ 18 years < 36

Reference range adopted from Abbott (vendor) based on in-house verification study of 20 normal female normal volunteers (>18 years old) in the UCSF Laboratory.

Pediatric Female Reference Range
Age Reference Range (U/mL)
0 - < 4 months 2 - 22
4 months - < 5 years 6-39
5 years - < 11 years 5 - 30
11 years - < 18 years 8-33

Pediatric reference ranges adopted from CALIPER Pediatric Reference Interval study performed on random samples using the Abbott Architect i2000 assay.

Reference range is not established for males.

Synonyms

  • CA125
  • C125

Storage/Transport Temperature

-20°C or colder

Stability (from collection to initiation)

Refrigerated (2-8°C):  7 days
Frozen (-20°C or colder): 12 months

If testing will be delayed more than 24 hours, remove serum from clot, red blood cells or serum separator gel.

Avoid multiple freeze-thaw cycles.

Reported

1-4 days

Additional Information

Method changed from Siemens Centaur XP platform to Abbott Architect i2000 platform on X3/19/2018. The Abbott Architect method reads approximately 30% higher than the Centaur method.  No changes were made to the reference range.

Results for this test are generated by immunoassay using the Abbott Architect i2000 platform.  Results for this test determined by assay methods from other manufacturers may not be comparable.

The ARCHITECT CA 125 II Calibrators are manufactured volumetrically and are referenced to this standard prepared by Fujirebio Diagnostics, Inc. There is no internationally recognized CA 125 standard available at this time.

CPT Codes

86304

LOINC Codes

10334-1