8 yellow-top (ACD) vacutainers of recipient blood and 1 gold-top (SST Gel) vacutainer of donor blood.
>20 mL of recipient blood and >6 mL of donor blood.
Additional Requirements
A drug treatment and sensitization history must be supplied which details dates of treatment with lymphocytotoxic agents (ATG, OKT3, rituximab, etc.) and/or sensitizing events (transplants, transfusions, and pregnancies).
Unacceptable Conditions
Broken tube, mismatched or missing labels, coagulated, >72 hours old, insufficient cellularity/volume, improper fixation, viability <50%, no history provided.
Storage/Transport Temperature
Room temperature. Do not freeze.
Stability (from collection to initiation)
72 hours at room temperature.
Performed
Monday through Friday from 9 A.M. to 5 P.M.
Reported
1-2 business days
Panel Components
Crossmatch reactivity.
Notes
Test is not in Cerner-Millennium. Please contact the lab at
(212) 305-6941.
Lab info
Pathology and Cell Biology
Section
Immunogenetics and Cellular Immunology
Methodology
Complement-Dependent Lymphocytotoxicity (CDC)
Reference Interval
No Reference Range
Can this test be added to a previous specimen?
No
CPT Codes
86805
Test Information
Patient Preparation
Routine venipuncture.
Collect
8 yellow-top (ACD) vacutainers of recipient blood and 1 gold-top (SST Gel) vacutainer of donor blood.
>20 mL of recipient blood and >6 mL of donor blood.
Additional Requirements
A drug treatment and sensitization history must be supplied which details dates of treatment with lymphocytotoxic agents (ATG, OKT3, rituximab, etc.) and/or sensitizing events (transplants, transfusions, and pregnancies).
Unacceptable Conditions
Broken tube, mismatched or missing labels, coagulated, >72 hours old, insufficient cellularity/volume, improper fixation, viability <50%, no history provided.
Storage/Transport Temperature
Room temperature. Do not freeze.
Stability (from collection to initiation)
72 hours at room temperature.
Performed
Monday through Friday from 9 A.M. to 5 P.M.
Reported
1-2 business days
Panel Components
Crossmatch reactivity.
Notes
Test is not in Cerner-Millennium. Please contact the lab at
(212) 305-6941.